• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION 2.0 LACTOSORB SYSTEM 2.0 X 5 MM DIRECT DRIVE SCREW; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION 2.0 LACTOSORB SYSTEM 2.0 X 5 MM DIRECT DRIVE SCREW; SCREW, FIXATION, BONE Back to Search Results
Model Number N/A
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00620, 0001032347-2020-00621, 0001032347-2020-00622.Concomitant medical products: 2.0 lactosorb system 2.0 x 5 mm direct drive screw, part# 915-2200, lot# 865090, 2.0 lactosorb system 2.0 x 7 mm direct drive screw, part# 915-2201, lot# 893700, 2.0 lactosorb system 2.0 x 5 mm direct drive screw, part# 915-2200, lot# 056430.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source: (b)(6).
 
Event or Problem Description
It was reported that the screws could not gripped during an oral surgery.Some of the screws were too tight and some of the screws were too loose.No adverse events have been reported as a result of the malfunction.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, g3, g6, h2, h3, h6, h10.The customer returned qty two 915-2200 lot 865090, qty two 915-2201 lot 893700, and qty two 915-2200 lot 056430 for evaluation due to lack of retention complaint.A visual inspection of the screw showed light signs of attempted use.For further analysis the screw was returned to warsaw for further investigation by the ops manufacturing technician.The reported testing results show that the returned screws were possibly damaged as there were imprints of the driver tip in the head of most the screws.Dimensional testing could not be completed due to the damage to the screw heads.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0 LACTOSORB SYSTEM 2.0 X 5 MM DIRECT DRIVE SCREW
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key11051030
Report Number0001032347-2020-00622
Device Sequence Number11422110
Product Code HWC
UDI-Device Identifier00841036054687
UDI-Public00841036054687
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K011139
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2025
Device Model NumberN/A
Device Catalogue Number915-2200
Device Lot Number056430
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/24/2020
Supplement Date Received by Manufacturer03/22/2021
Initial Report FDA Received Date12/21/2020
Supplement Report FDA Received Date03/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-