| Model Number |
1365-60-000 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/09/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported as metal on metal hip that mri showed effusion.Soft tissues were stained with metal debris.Pinnacle metal liner explanted and replaced with poly liner.Patient had elevated cobalt levels.Doi: unknown.Dor: (b)(6) 2020.Affected side: left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Review of the singular provided x-ray image finds nothing indicative of a product problem.The image does not aid in determining a root cause for the reported allegations.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : per wi-3430 it has been determined that should related reports be identified an mre is not required.Corrected h6 health effect - clinical code: appropriate term / code not available (e2402) from the initial medwatch to capture blood heavy metal increased.
|
| |
|
Search Alerts/Recalls
|