• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE RX PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE RX PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U41502H30RX
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 03/2023); device pending return.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon was allegedly twisted.Reportedly, it was difficult to remove the balloon and the catheter shaft allegedly detached during removal.The distal catheter segment was removed using guiding sheath.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one ultraverse rx catheter in two segments has returned for evaluation.On the visual evaluation of the device, it appeared to be bloody and the device returned in two segments in which one part contains the inflation hub catheter and partial inflation lumen.The second segment contains remaining inner lumen with the distal tip inverted within the balloon.No other specifics anomalies noted.No functional performed due to the condition of the device.Therefore, the investigation as confirmed for the reported detachment of device has confirmed as the device returned in two segments.The investigation for reported difficult to remove remains inconclusive as no evidence noted to the device which returned for evaluation.The investigation remains inconclusive for the reported material twist as no evidence identified in the evaluation of returned device.Although it is likely the alleged difficult to remove led to the alleged detachment of device.A definitive root cause for the alleged material twist, detachment of device and difficult to remove could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 03/2023), g3.H11: h6 (method, result and conclusion).H11:section a through f the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the lesion below the knee, the pta balloon was allegedly twisted.Reportedly, it was difficult to remove the balloon and the catheter shaft allegedly detached during removal.The distal catheter segment was removed using guiding sheath.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE RX PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11054088
Report Number2020394-2020-20906
Device Sequence Number19472967
Product Code LIT
UDI-Device Identifier00801741056710
UDI-Public(01)00801741056710
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K131199
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU41502H30RX
Device Catalogue NumberU41502H30RX
Device Lot NumberCMER0291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Initial Date Received by Manufacturer 11/30/2020
Supplement Date Received by Manufacturer05/23/2021
Initial Report FDA Received Date12/22/2020
Supplement Report FDA Received Date05/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-