B. BRAUN MEDICAL INC. APEX; COMPOUNDER Back to Search Results
Model Number AX1000
Medical Device Problem Code Application Program Problem: Dose Calculation Error (1189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number: (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As per reported by the user facility: the solution on line 20 changed from sodium phosphate to potassium phosphate in the apex system.There was no patient involvement.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The logs were provided for evaluation.The logs show that sodium phosphate was confirmed on station 20 and later a potassium phosphate ndc was scanned and confirmed on station 20.It was found that the two (2) solutions were marked as interchangeable in the database.Apex will allow solutions that are marked interchangeable to be switched on the same station throughout the day and when reconciling an order would allow the ordered volume to be filled by either interchangeable ingredient.To fix this, the user needs to go into either the sodium phosphate ingredient or the potassium phosphate ingredient and undo the interchangeability between the two.If additional pertinent information becomes available a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APEX
Common Device Name
COMPOUNDER
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
1601 wallace drive, suite 150
carrollton TX 75006 6690
MDR Report Key11054671
Report Number1641965-2020-00024
Device Sequence Number2602478
Product Code NEP
UDI-Device Identifier04046955048502
UDI-Public04046955048502
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAX1000
Device Catalogue NumberAX1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/10/2020
Supplement Date Received by Manufacturer12/10/2020
Initial Report FDA Received Date12/22/2020
Supplement Report FDA Received Date01/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-