| Model Number |
620RG25 |
| Medical Device Problem Codes |
Material Separation (1562); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
Mitral Valve Stenosis (1965)
|
| Date of Event |
11/18/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that 5 years and 20 days post implant of this 25mm mitral annuloplasty ring, it was explanted and replaced with a 27mm mitral bioprosthetic valve.It was reported that the ring had become too small for the patient and was causing mitral stenosis.It was also reported that the anterior annular part of the ring was detached from the mitral annulus.No additional adverse patient effects were reported. .
|
| |
|