| Model Number |
1987-16-212 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Fall (1848); Inadequate Osseointegration (2646)
|
| Date of Event |
12/09/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a distal femoral replacement on (b)(6) 2020.He suffered a fall with subsidence of the femur and loosening of the porous stem (non-cemented) at the bone to implant interface.The patient was brought to surgery and the stem was replaced with a larger stem.Additional segmental portions were placed and the distal femur implant was reused.Then, the tibial insert was placed.No surgical delay was noted.Doi: (b)(6) 2020, dor: (b)(6) 2020, right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|