ST. JUDE MEDICAL AGILIS NXT STEERABLE INTRODUCER DUAL-REACH 61 CM SHEATH, SMALL CURL DUAL-REACH; INTRODUCER, CATHETER Back to Search Results
Model Number G408318
Device Problem Fluid/Blood Leak (1250)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During the procedure, a blood leak was noted from the handle part of the introducer.The procedure was completed without replacing the device and there were no adverse consequences to the patient.
 
Manufacturer Narrative
One 8.5f agilis steerable introducer sheath was received for evaluation.A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal seal.There was no evidence of a leak within the handle.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seal and subsequent leak is consistent with damage during use.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILIS NXT STEERABLE INTRODUCER DUAL-REACH 61 CM SHEATH, SMALL CURL DUAL-REACH
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica MN 1897- 4050
CS  1897-4050
MDR Report Key11056997
MDR Text Key223182922
Report Number3005334138-2020-00644
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734206112
UDI-Public05414734206112
Combination Product (y/n)N
PMA/PMN Number
K061363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model NumberG408318
Device Catalogue NumberG408318
Device Lot Number7635230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2021
Was the Report Sent to FDA? No
Date Manufacturer Received01/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-