| Model Number |
66021306 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Skin Tears (2516)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, the allevyn life sacrum is "shedding" silicone onto the patient upon removal and it's difficult to remove from the skin.It's also shedding as the backing is removed before application.It is unknown how the procedure was finished or if there was any delay.
|
| |
|
Additional Manufacturer Narrative
|
|
We have now concluded our investigation for the complaint received.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.The device used in treatment has not been returned for evaluation, with no additional information provided.We have not been able to establish a relationship between the device and the reported event or determine a root cause on this occasion.Potential factors include raw material issues or storage environment issues.The complaint history file contains further instances of the reported event, currently being monitored, with actions being taken to reduce further instances.This investigation is now complete with no further action deemed necessary at this stage.Smith + nephew are taking further actions relating to the failure reported and continue to monitor for adverse trends.
|
| |
|
Search Alerts/Recalls
|