| Model Number |
1307-61-142 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
12/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The patient was revised to address shoulder dislocation.Reason for dislocation is undetermined as per surgeon.No delay to surgery reported.No additional information available.Doi: (b)(6) 2020, dor: (b)(6) 2020, left shoulder.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|