• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MBT REVISION REAMER 10MM; KNEE INSTRUMENT : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MBT REVISION REAMER 10MM; KNEE INSTRUMENT : REAMERS Back to Search Results
Model Number 2178-63-170
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: examination of the returned device confirmed the reported event.The investigation attributed the root cause to normal use and servicing and did not establish that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event or Problem Description
It was reported that all reamers were loose on the adapters and spun freely.Adapter would spin freely when attached to reamers or any other attachment.T-handle was also loose.Attachment point seems to be weak and came free at first sign of resistance.No pieces broke off in patient.All pieces were removed from patient.Surgery time was extended by approximately 30 as a result of depuy instruments not functioning properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT REVISION REAMER 10MM
Common Device Name
KNEE INSTRUMENT : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11057538
Report Number1818910-2020-27543
Device Sequence Number2651360
Product Code HTO
UDI-Device Identifier10603295096276
UDI-Public10603295096276
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2178-63-170
Device Catalogue Number217863170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/15/2020
Initial Report FDA Received Date12/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-