Catalog Number 71932634
Medical Device Problem Code Device Fell (4014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during a tka procedure, the gen ii 7.5mm resur pat 26mm patella fell off of the cement that had adhered to the patient's native patella.The procedure was successfully completed using a back-up device.It is unknown if there was any delay in the procedure.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not received for evaluation and the reported event could not be confirmed.All efforts were made to locate the device but we have no evidence that it was ever received by the complaint investigation team for evaluation.The clinical/medical investigation concluded that, this case reports that after removing the patella cement clamp, the patella completely fell off of the cement that had adhered to the patient's native patella.Per the sales rep, a new/backup patella was opened and implanted.There was no patient harm reported, however it is unknown whether there was surgical delay.Since no patient harm is alleged, no further clinical assessment is warranted.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible probable cause could include but not limited to procedural variance, product preparation or storage condition.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEN II 7.5MM RESUR PAT 26MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11057980
Report Number1020279-2020-07717
Device Sequence Number3310833
Product Code JWH
UDI-Device Identifier00885556056233
UDI-Public885556056233
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71932634
Device Lot NumberN/A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/01/2020
Supplement Date Received by Manufacturer08/20/2021
Initial Report FDA Received Date12/22/2020
Supplement Report FDA Received Date08/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-