• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 5CM HEAVY DUTY SHORT ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 5CM HEAVY DUTY SHORT ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) Back to Search Results
Model Number SHORT-HD
Medical Device Problem Codes Vibration (1674); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
 
Event or Problem Description
It was reported by netherlands that during service and evaluation, it was determined that the ball bearings and bearing of the attachment device were worn, and the triangle symbol was missing.It was determined that the device generated heat and vibration, and the labeling was illegible and etching.It was further determined that the device failed pretest for labeling assessment, vibration and temperature.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5CM HEAVY DUTY SHORT ATTACHMENT
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key11058381
Report Number1045834-2020-02181
Device Sequence Number3206752
Product Code ERL
UDI-Device Identifier00845384008804
UDI-Public00845384008804
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K011444
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSHORT-HD
Device Catalogue NumberSHORT-HD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2020
Initial Report FDA Received Date12/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-