| Model Number |
1650 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Additional information has been requested regarding the patient's age, but it was not available at the time of this report.If additional information is received, the event will be updated and a supplemental report will be sent.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the battery exhibited power switching.The issue occurred when connected to either side of the controller.After about five minutes, the controller was switching over to the other battery even when the controller stated the battery was three-quarters to one hundred percent full.The battery will be replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for device evaluation.Product event summary: the battery was not returned for evaluation.Analysis of the data log file revealed several premature power switching events that were due to momentary disconnections involving the battery.As a result, the reported event was confirmed.A power source lubrication procedure was performed in accordance with the requirements under fsca cvg-18-q4-19 on (b)(6) 2018 to mitigate the reported conditions and the battery remained in use.The most likely root cause of the reported event can be attributed to momentary disconnections due to temporary corrosion of the controller-port/power-source pins.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|