(b)(4).Events reported in mwr- 2210968-2020-10215, 2210968-2020-10217.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
(b)(4).Date sent to the fda: 3/2/2021.A manufacturing record evaluation was performed for the finished device batch qbbptk and no non-conformances were identified.H6 component code: g07002 ¿ conforming device.Additional h3 investigation summary: visual analysis of the returned sample revealed that the 8762h sample was returned, and swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along of the strand and no defects or damage were observed.A functional test was performed and the pull force was above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, the device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The device performed without any defect noted.Two sealed boxes (36 ea.) and an opened box with fourteen unopened samples were returned to ethicon inc for evaluation.Visual inspection and functional testing were conducted on thirteen returned devices.Visual analysis of the returned samples revealed that the 8762h samples were returned with no apparent damage on the packages.The samples were opened, and swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|