| Model Number |
3100A |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).A vyaire field service representative (fsr) went onsite and evaluated the ventilator.It was seen that the mean airway pressure (map) panel meter connection was loose.Map display is bouncing, and won't settle.Fsr reseated the map panel meter connection, and the map reading increased and settled.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported to vyaire medical that the 3100a ventilator has a leak heard from the regulators in the pneumatics.At this time, there is no information regarding patient involvement associated with the reported event.
|
| |
|