A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp and was unable to reproduce the reported issue.However, based on the fault codes in the unit's logs, the fse replaced the executive processor pcb.Unrelated to said reported issue, the fse also found the fiber optic connector and p-clip were damaged and replaced same, respectively.Moreover, the fse replaced both expired lithium-ion batteries, and replaced the expired critical alarm 9v battery.After repairs were completed, the unit was left to run over night while connected to a system trainer and test balloon with no failures.The fse then performed a full calibration, all functional and safety checks to meet factory specifications.The iabp was then released to the customer for return to clinical service.(b)(6).
|