The user facility reported that the involved capiox device was used during the procedure.
Bsa 1.
72.
Blood gas test result with cdi500, 50 minutes after pump started: blood flow 4.
2l, gas flow 2.
5l, fio2 60%, pao2 around 250, and co2 around 50.
After that, co2 increased to 80 while pao2 did not change.
In order to decrease co2, another fx25 was added to return blood to the reservoir, however it could not be decreased as expected (from 80 to 70 only).
After the first oxy was removed, circulation continued with the second oxy only, and pao2 and co2 became stable.
No pressure drop was observed during ecc.
Second fx25 was added to the arterial side.
Oxygenated blood was returned to the reservoir and then pumped through the first oxy.
After that, the first oxy was removed and circulation was done with the second oxy only.
The patient was not harmed.
The procedure outcome was not reported.
|