• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY, WIRELESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY, WIRELESS Back to Search Results
Catalog Number 04P75-01
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2020, abbott point of care (apoc) was contacted by a customer who reported that there was a noise and smoke came from the battery terminals of i-stat 1 analyzer sn# (b)(4).There was no additional information at the time of this report.The analyzer was replaced at no charge and returning for investigation.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the power source is unknown at the time of this report.The product was replaced and returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Event or Problem Description
Na.
 
Additional Manufacturer Narrative
Apoc incident: (b)(4).The investigation was completed on (b)(6) 2021.The customer reported that analyzer s/n (b)(6) made a "bang" and smoke emanated from the battery compartment.There is insufficient evidence to determine the cause of the complaint as the analyzer has not been returned to flex.A rocketware search spanning six months revealed no similar incidents and no evidence of a trend.No deficiency has been identified.
 
Additional Manufacturer Narrative
Apoc incident: (b)(4).The investigation was completed on (b)(6) 2021.The customer reported that analyzer s/n (b)(6) emitted a "bang" sound and smoke emanated from the battery compartment.Failure analysis did not reproduce either part of the complaint and the analyzer functioned according to specification throughout testing.The batteries that were in use at the time of the incident were not returned, and therefore they cannot be ruled out as a contributing factor.A rocketware search spanning one year revealed one similar incident and no evidence of a trend.No deficiency has been identified.
 
Event or Problem Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY, WIRELESS
Common Device Name
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key11064402
Report Number2245578-2020-00135
Device Sequence Number2560300
Product Code CGA
UDI-Device Identifier00054749000340
UDI-Public00054749000340
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K001387
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04P75-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2021
Initial Date Received by Manufacturer 12/14/2020
Supplement Date Received by Manufacturer03/01/2021
10/11/2021
Initial Report FDA Received Date12/23/2020
Supplement Report FDA Received Date03/02/2021
10/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-