Model Number A701125
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned due to a temperature issue.The temperature accuracy read as 192.1°c and was not within specification.Microscopic analysis revealed char on the distal tip.Further investigation revealed it is possible the high temperature is related to incorrect length of the pi tubing; however, it could not be confirmed due to the adhesive around the pi tubing.
 
Event or Problem Description
This report is to advise of an event observed during analysis confirming a high ablation temperature noted while testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXABILITY ABLATION CATHETER, BI-D, CURVE D-F
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key11064511
Report Number3008452825-2020-00719
Device Sequence Number2876584
Product Code OAD
UDI-Device Identifier05415067011114
UDI-Public05415067011114
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P110016
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Model NumberA701125
Device Catalogue NumberA701125
Device Lot Number6822866
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2020
Initial Report FDA Received Date12/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
Patient Weight90
-
-