• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number VPR-GW-14
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The guide wire was received at csi for analysis.Visual examination revealed the guide wire was fractured, and that the spring tip was intact with no damage observed.Scanning electron microscopy analysis revealed evidence of fatigue and a large crack down the middle of the guide wire, which was consistent with a fracture due to use of the wire in an unusually elevated stress environment.At the conclusion of the device analysis investigation, the report that the guide wire had fractured was confirmed, however the exact root cause of the fractures could not be determined.The material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
 
Event or Problem Description
Orbital atherectomy treatment was successfully performed in the superficial femoral artery, and the peripheral orbital atherectomy device was removed from the patient.Multiple balloons were inserted over the viperwire guide wire and balloon angioplasty was performed.The balloons and wire were removed from the patient.An angiogram was performed and revealed that part of the guide wire had fractured and remained in the sheath.The fragment was retrieved from the sheath with a snare device.The patient condition following the procedure was described as "fine.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
MDR Report Key11064551
Report Number3004742232-2020-00417
Device Sequence Number19472917
Product Code MCW
UDI-Device Identifier10852528005077
UDI-Public(01)10852528005077(17)220430(10)328047
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model NumberVPR-GW-14
Device Catalogue Number72023-01
Device Lot Number328047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2020
Initial Report FDA Received Date12/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-