HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Catalog Number 425-00
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2020
Event Type  malfunction  
Manufacturer Narrative
Qn #: (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A verification of the reported failure mode was conducted and 8 devices were taken from current production (425-00 concha neptune lot # 73l200012n) at the manufacturing facility and were functionally tested.No issues were encountered during the functional testing.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported that the device "overheats then turns off".No patient involvement reported.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was prepared for the functional bench test where water, an adult breathing circuit (880-36kit), 2-3 lpm of air pressure, and a 382-10 concha smart water column was connected to the unit for a real time operational scenario.During the pre-operational self-tests a red wench appeared and the unit alarmed.The unit was opened and the resistance was measured across the thermal switch and the thermal fuse.The resistance across the switch measured 0.2 ohms, which is within the normal operating range of 0.2-0.4 ohms.The switch measured 0.1 ohms which confirms it is not blown.Based on the above, the only other component that would cause a red wrench failure is the power control board.This could have resulted from a faulty component on the board or a loose solder connection.Based on the investigation performed, the reported complaint is confirmed.Functional testing revealed a faulty power control board.It is unknown what caused this control board to fail.It could have been a faulty component on the board, or a solder joint got loose during shipping and handling.Based on the sample returned and the complaint description the root cause cannot be determined.
 
Event Description
It was reported that the device "overheats then turns off".No patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
MDR Report Key11064953
MDR Text Key223396806
Report Number3003898360-2020-01007
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Date Manufacturer Received01/14/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-