| Model Number |
07P9920 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.There was no additional patient information provided by the customer due to privacy issues.
|
| |
|
Event or Problem Description
|
|
The customer reported a (b)(6) alinity c creatinine result on one male patient.The results provided were on (b)(6) 2020 initial = 400umol/l (normal is 110) / retest the next day on a new sample, same patient = 73umol/l.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of tickets by lot 26956un20 found no other complaints for this issue.Trending review determined no trends for elevated patient results due to the product.Return testing was not completed as returns were not available.Device history record review did not identify any non-conformances or deviations for lot 26956un20 and the complaint issue.A review of product labeling concluded that the issue is sufficiently addressed.Based on the investigation, no systemic issue or deficiency of alinity c creatinine reagent, lot number 26956un20 was identified.
|
| |
|