| Model Number |
TRANSDUCER X7-2 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Mechanical Jam (2983)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported an x7-2t model transducer had an articulation issue during use.There was no injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A thorough evaluation of the suspect transducer could not reproduce the articulation issue as described by the customer.The transducer passed all performance testing, however, some cosmetic damage was identified.The investigation of the device revealed bite damage and scratches to the beading, window, and tipshell of the transducer.The physical damage to the transducer is indicative of improper maintenance.Although there was evidence of cosmetic damage, the investigation confirmed the transducer was operating as designed with no malfunctions.
|
| |
|