| Model Number |
1012448-15 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use and there was no report of a leak during preparation for use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the very tortuous anatomy and/or other devices resulted in compromising the balloon material; thus resulting in the reported material rupture.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion located in the tortuous distal right coronary artery (rca).The 2.75 x 15 mm nc trek balloon ruptured.The device was replaced with a new balloon catheter to complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.
|
| |
|