• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012447-15
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2020
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that tip of a 2.5x15mm nc trek balloon dilatation catheter (bdc) separated in the introducer guide.All product was retrieved as a unit.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Visual inspection was performed on the returned device.The reported separation was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints from this lot.The investigation determined the reported separation appears to be related to operational context there is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11067736
MDR Text Key223600027
Report Number2024168-2020-10917
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648151767
UDI-Public08717648151767
Combination Product (y/n)N
PMA/PMN Number
K103153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Model Number1012447-15
Device Catalogue Number1012447-15
Device Lot Number90613G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Date Manufacturer Received02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age78 YR
-
-