| Model Number |
DBP-EX-150SOL145 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id:(b)(4).
|
| |
|
Event or Problem Description
|
|
During orbital atherectomy treatment on a lesion in the popliteal artery, the peripheral exchangeable orbital atherectomy device (oad) stalled and stopped spinning.The oas pump light was observed to be yellow and all connections were found to be secure.The pump was powered on and all leds on the oad handle were illuminated, so the device was removed.During device removal the oad was unable to move backward on the guide wire and both the wire and device were removed from the patient.The lesion was re-wired and the procedure was completed with a second oad.Due to the time required to remove the device and re-wire the lesion, a delay of 30 minutes or greater occurred.There were no patient adverse events.
|
| |
|