Model Number N/A |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/30/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The getinge field service engineer (fse) that encountered the issue replaced the batteries and completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery run time test.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery run time test.There was no patient involvement, and no adverse event reported.
|
|