| Model Number |
482802740 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately 3 weeks following the initial surgery, an es2 integrated blade screw and mantis straight rod fractured post-operatively.Revision surgery has been performed.This record represents the blade screw.
|
| |
|
Additional Manufacturer Narrative
|
|
The screw was returned with the tulip disengaged from the screw shank.The tulip locking ring was deformed.The shank bulb had a deep and angled rod indentation, indicative of tulip angulation and overtightening.There is also a sharp angulation midway through the shank bulb indentation, indicating that rod tightening was performed at least twice at tulip head/rod over angulation and tulip head/rod severe over angulations.It was stated bone fusion was not achieved.Device and complaint history records were reviewed for this lot, and no relevant manufacturing issues or similar complaints were identified.It is unknown what torque the blockers were tightened to, the surgical technique states they must be tightened to 12nm and anti torque must be used.It is unknown if anti torque was used.Based on the deformation found on the returned screw, the most likely cause is multi factorial.Possible causes for tulip disengagement include overtightening of the blocker, fractured rod causing an unstable construct, excessive rod angulation and/or excessive post op activity.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 3 weeks following the initial surgery, an es2 integrated blade screw and mantis straight rod fractured post-operatively.Revision surgery has been performed.This record represents the blade screw.
|
| |
|
Search Alerts/Recalls
|