• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II XR INS XLPE LAT 3-4 RT 8MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CNSTRND, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II XR INS XLPE LAT 3-4 RT 8MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CNSTRND, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 74023830
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 12/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that a primary journey xr knee replacement was done by another surgeon in (b)(6) 2020.Then, the patient was revised on (b)(6) because the patient complained of knee pain and could not flex the knee past 95 degrees after the initial operation.The surgeons identified that the problem was because the joint line was raised by about 10 mm and the tibial slope was decreased from about 10 degrees to zero.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that, this case reports a journey ii revision was performed due to pain and decreased flexion.Per email communication, there is no additional information available to fully evaluate the root cause of the reported event.Without clinically relevant patient-specific supporting documentation a thorough medical investigation cannot be performed.No further medical assessment is warranted at this time.Should clinically relevant documentation/ information become available, the clinical/medical task may be re-evaluated.A review of complaint history did not reveal additional complaints for the listed device.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes could be alignment, fit/size of device used or wear.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II XR INS XLPE LAT 3-4 RT 8MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CNSTRND, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11069704
Report Number1020279-2020-07754
Device Sequence Number14254830
Product Code JWH
UDI-Device Identifier00885556615607
UDI-Public00885556615607
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K152726
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74023830
Device Catalogue Number74023830
Device Lot Number18ESL0015
Initial Date Received by Manufacturer 12/01/2020
Supplement Date Received by Manufacturer07/19/2021
Initial Report FDA Received Date12/23/2020
Supplement Report FDA Received Date07/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-