Model Number 3CX*NX19RW
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo (b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, the clinician noticed a leak at the inlet while priming.*no patient involvement *product was changed out *procedure completed successfully.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on december 18, 2020.  upon further investigation of the reported event, the following information is new and/or changed: h6 (adverse event problem 527, 2645, 4582, 67, 11).A second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: the affected sample was inspected upon receipt to note slight damage to the venous inlet of the reservoir and the blood inlet of the oxygenator.The affected sample was setup to circulate water and observe for leakage.A representative retention sample was inspected with no visual damage.It then was setup to circulate water for leakage.No leakage were observed on the actual and retention samples.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on december 18, 2020.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 527, 4582).Component code: 527 - valve.Health effect clinical code: 4582 - no clinical signs, symptoms or conditions.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE NX19RW W/4L RESERVOIR
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key11070220
Report Number1124841-2020-00283
Device Sequence Number2667198
Product Code DTZ
UDI-Device Identifier00699753450875
UDI-Public(01)00699753450875
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K180950
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model Number3CX*NX19RW
Device Catalogue NumberN/A
Device Lot NumberXP02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/01/2020
Supplement Date Received by Manufacturer01/06/2021
01/28/2021
07/02/2021
Initial Report FDA Received Date12/23/2020
Supplement Report FDA Received Date01/26/2021
02/03/2021
07/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-