• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII MI IN SLT 3 MO TI CU BL RT; PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII MI IN SLT 3 MO TI CU BL RT; PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,METAL/POLY Back to Search Results
Model Number 71441137
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Case- (b)(4).
 
Event or Problem Description
It was reported that, during a training the tibial cutting block screw broke.No case involved; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was not returned for evaluation.Therefore, a product analysis could not be performed at this time and the reported event could not be confirmed.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MI IN SLT 3 MO TI CU BL RT
Common Device Name
PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11072107
Report Number1020279-2020-07772
Device Sequence Number14254833
Product Code HRY
UDI-Device Identifier03596010484093
UDI-Public03596010484093
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number71441137
Device Catalogue Number71441137
Device Lot Number06CM08410
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/01/2020
Supplement Date Received by Manufacturer04/26/2021
Initial Report FDA Received Date12/24/2020
Supplement Report FDA Received Date05/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-