Catalog Number 8065751766
Device Problems Positioning Failure (1158); Dull, Blunt (2407)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that the trocar was not sharp and was not possible to place in the sclera.The product was replaced and the procedure was completed.There was no patient harm.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of the trocars are not sharp enough; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Because a sample was not received at the manufacturing site and the device history record review indicated that the product was processed and released according to the product¿s acceptance criteria, the exact root cause for this complaint is unknown.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All trocar blades are 100% inspected.Non-conforming product is removed from the lot.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The previous report submitted for this event contained an error, in h.1.¿summary report¿ was inadvertently selected.After a recent systems update, a system error caused the inadvertent additional selection of ¿summary report¿ in h.1.On a select number of reports.The error, which was limited only to the h.1.Field, was promptly identified and quickly rectified.Correction smdrs are being filed for the impacted reports.This smdr is correcting that error for this event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key11072177
MDR Text Key224414431
Report Number1644019-2020-00724
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Catalogue Number8065751766
Device Lot Number2392632H
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/04/2020
Initial Date FDA Received12/24/2020
Supplement Dates Manufacturer Received02/25/2021
03/03/2021
Supplement Dates FDA Received03/04/2021
03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No