MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND Back to Search Results
Model Number 1000354
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2020
Event Type  malfunction  
Manufacturer Narrative
The user guide states: your t:slim pump can stop insulin delivery and alert you (or whoever is with you) if there has been no interaction with the pump within a specified period of time.The default for this alarm is preset to 12 hours.You can set it anywhere between 5 and 24 hours, or off.This alarm notifies you that there has been no interaction with the pump in the specified number of hours and the pump will shut down after 30 seconds.No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the customer received an auto-off alarm.There was no adverse impact to the customer¿s blood glucose value.Tandem technical support educated the customer on the pump's auto-off safety feature per user guide and advised the customer to consult with healthcare provider prior to modifying the setting.Multiple attempts were made by tandem technical support to obtain additional information; however, no information was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Type of Device
AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key11072183
MDR Text Key223606041
Report Number3013756811-2020-150328
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00853052007325
UDI-Public(01)00853052007325
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201214
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 12/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000354
Device Catalogue Number1000912
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 12/14/2020
Initial Date FDA Received12/24/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age52 YR
-
-