• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7010
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 35% stenosed target lesion was located in the middle of left anterior descending artery.A 2.75mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, during inflation upon reaching the rated pressure, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient's status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 35% stenosed target lesion was located in the middle of left anterior descending artery.A 2.75mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, during inflation upon reaching the rated pressure, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient's status was stable.It was further reported that the lesion was mildly tortuous and severely calcified.The balloon was inflated once and ruptured at 15 atmospheres for 25 seconds.The balloon was removed within the catheter.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 35% stenosed target lesion was located in the middle of left anterior descending artery.A 2.75mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, during inflation upon reaching the rated pressure, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient's status was stable.It was further reported that the lesion was mildly tortuous and severely calcified.The balloon was inflated once and ruptured at 15 atmospheres for 25 seconds.The balloon was removed within the catheter.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.The balloon was tightly folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There were no issues detected.Inspection of the remainder of the device presented no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11072283
Report Number2134265-2020-18457
Device Sequence Number11460283
Product Code LOX
UDI-Device Identifier08714729392668
UDI-Public08714729392668
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2023
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0025288199
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Initial Date Received by Manufacturer 12/18/2020
Supplement Date Received by Manufacturer12/24/2020
02/11/2021
Initial Report FDA Received Date12/24/2020
Supplement Report FDA Received Date01/11/2021
02/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight60
-
-