| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device would not power up.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation.The customer's report was not replicated or confirmed.The device passed all testing including full functionality testing and stress testing without duplicating the report.Review of the device log shows no power-related entries that would suggest the device did not power up via battery.There is no evidence of low battery, shutdown warnings, or device resets.The device was recertified and returned to the zoll loaner inventory.Analysis of reports of this type has not identified an increase in trend.
|
| |
|