| Model Number |
1012534-20 |
| Medical Device Problem Codes |
Break (1069); Difficult to Insert (1316); Material Twisted/Bent (2981); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the absolute pro stent delivery system was unable to be inserted/back loaded unto the unspecified guide wire.A bending was then observed on the absolute pro catheter.The exact location of the bend was not specified by the account.The device was set aside and not used.There was no adverse patient effect or other device issues reported.Another device was successfully used in replacement.No additional information was provided regarding this issue.Subsequently the device was received at abbott vascular and preliminary analysis revealed multiple tears on the jacket at the location of the kink.The inner braiding material was separated (not in two separate pieces) within the jacket at the location of the kink.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual and functional analysis was performed on the returned device.The reported difficulty advancing the guide wire was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints.The investigation determined that the reported difficulties were likely related to procedural circumstances.It is likely that the difficulty encountered advancing the guide wire was the result of the severe kink/tear noted on the returned unit.Although a cause for the damage could not be determined, it may be possible that inadvertent mishandling during removal from the packaging or during preparation caused shaft to kink preventing the guide wire from advancing.Additional handling and the continued attempt to advance the guide wire with the shaft kinked may have caused the shaft to tear preventing any further advancement of the guide wire.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|