| Model Number |
CD3265-40 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported the asymptomatic patient presented remotely.Upon review of the transmission, it was observed the implantable cardioverter defibrillator was post paced t-wave oversensing.No programming changes were completed to address this issue.There were no patient consequences.
|
| |
|
Event or Problem Description
|
|
New information received indicated the recommended programming changes were completed to address the issue.The patient was stable.
|
| |
|