ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012622-39
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device code (b)(4).The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported the ominilink stent delivery system was prepped before removal of the protective sheath, and the stent came off the balloon.The device was not used, and there was no patient involvement.There was no clinically significant delay.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported stent dislodgment was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints from this lot.The device was prepped prior to removal of the protective cover.The omnilink elite instruction for use instructs to remove the protective cover from the tip prior to stent delivery system preparation.Based on the information provided, the stent dislodgment was the result of preparing the device prior to removal of the protective sheath.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11074128
Report Number2024168-2020-10941
Device Sequence Number14128858
Product Code NIO
UDI-Device Identifier08717648179204
UDI-Public08717648179204
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number1012622-39
Device Catalogue Number1012622-39
Device Lot Number8120341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Initial Date Received by Manufacturer 12/07/2020
Supplement Date Received by Manufacturer01/13/2021
Initial Report FDA Received Date12/27/2020
Supplement Report FDA Received Date01/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-