| Model Number |
1012622-39 |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device code (b)(4).The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported the ominilink stent delivery system was prepped before removal of the protective sheath, and the stent came off the balloon.The device was not used, and there was no patient involvement.There was no clinically significant delay.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported stent dislodgment was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints from this lot.The device was prepped prior to removal of the protective cover.The omnilink elite instruction for use instructs to remove the protective cover from the tip prior to stent delivery system preparation.Based on the information provided, the stent dislodgment was the result of preparing the device prior to removal of the protective sheath.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
|
| |
|