| Model Number |
2P56-22 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.There was no additional patient information provided by the customer due to privacy issues.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated ldh results on one patient generated on the architect analyzer.The results provided were: on (b)(6) 2020 (b)(6) initial ldh on serial number (b)(4) (normal range 125-220u/l) / retest on sn (b)(4).There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of similar complaint tickets could not be performed as the ldh reagent lot number was unknown.Trending review determined no trends for falsely elevated results for the product.Return testing was not completed as returns were not available.The patient sample was retested on the same instrument and acceptable results were generated.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation, no systemic issue or deficiency of the lactate dehydrogenase (ldh) reagent, list number 2p56 was identified.
|
| |
|