| Model Number |
705020 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
12/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "surgeon attempted to remove a locking screw with manual screwdriver from axsos distal femur plate.The tip of the screwdriver broke off in the patient from excessive force.A small fragment of the screwdriver had to be left in the patient.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the device was returned for evaluation and matches the reported failure mode.The device inspection revealed the following: the received screwdriver bit shows signs of normal and prolonged usage however its tip is found to be broken.The breakage surface shows signs of slight rotational deformation/waves.Absence of plastic deformation indicates that the breakage was instantaneous (brittle) due to high torsional and bending overload.However, it cannot be excluded that residual stress from previous usage also contributed in the breakage.The critical diameter of the drill was observed to be 5.29mm as required against a range of 5.20mm to 5.40mm.Similarly, the average hardness of the drill was observed to be 60.8 hrc as required against a range of 58.0 hrc ¿ 62.0 hrc.As per the dimensional inspection and hardness testing, the returned drill was found to be within specifications.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the failure is user related.The breakage of the tip happened due to a torsional and bending overload, the likelihood of which is greater during an explantation.If any further information is provided, the complaint report will be updated.
|
| |
|
Event or Problem Description
|
|
As reported: "surgeon attempted to remove a locking screw with manual screwdriver from axsos distal femur plate.The tip of the screwdriver broke off in the patient from excessive force.A small fragment of the screwdriver had to be left in the patient.".
|
| |
|