| Model Number |
KD-650Q |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bowel Perforation (2668)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device referenced in this report has not been returned to olympus for evaluation.The definitive cause of the customer's experience cannot be determined at this time.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.This reported event has been reported by the importer on mdr# 2951238 - 2020 - 00529.
|
| |
|
Event or Problem Description
|
|
It is reported in the literature article titled "esd with double-balloon endoluminal intervention platform versus standard esd for management of colon polyps" appearing in the aug 2020 issue of endoscopy international, following a colonoscopy with double balloon endoluminal intervention using a pentax single-channel video endoscope, an olympus disposable distal cap, and an olympus single use electrosurgical knife, one patient experienced a bowel perforation.No information was provided regarding any treatment required as a result of this occurrence.There was no report of any olympus device malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being updated to provide additional investigation findings.A device history record (dhr) review for this specific device could not be carried since no lot number was reported.Olympus does not ship any device that doesn't meet all design and quality specifications.Conclusion: there is no indication that the device malfunctioned.The reported event is an anticipated potential adverse outcome of the stated procedure.
|
| |
|
Search Alerts/Recalls
|