• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIMING ARM FOR SUPRAPATELLAR; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIMING ARM FOR SUPRAPATELLAR; NAIL, FIXATION, BONE Back to Search Results
Model Number 03.010.441
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, during an unknown procedure, while drilling, the surgeon hit the nail and devices (insertion handle, cannulated screw, protection sleeve, drill sleeve, aiming arm for supratellar, titanium cannulated tibial nail, and drill bit) did not line up completely.The procedure was successfully completed with no surgical delay.No patient consequences.This complaint involves seven (7) devices.This is report 5 of 7 for (b)(4).
 
Event or Problem Description
Concomitant devices reported: nail (part unknown, lot unknown, quantity 1).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part 03.010.441, lot 180064-103: manufacturing site: selzach.Supplier: (b)(4).Release to warehouse date: may 23, 2018.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.H3, h6: a product investigation was completed: upon visual inspection it was noticed that not all the reported devices were returned for possible functional verification.There were no defects noticed on the device itself.Not all the devices were returned, but the returned devices were put together and there was no issue in the interaction of the devices.The alignment cannot be verified with the partial devices returned and no complete information is available.There were no issues identified with the returned device that could have caused the complaint condition, so the dimensional inspection was not performed.Based on the date of manufacture, the current and manufactured revision of drawings were reviewed.The complaint condition is not confirmed for the returned part.No definitive root cause could be determined based on the provided information.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING ARM FOR SUPRAPATELLAR
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11078246
Report Number2939274-2020-05727
Device Sequence Number11506962
Product Code JDS
UDI-Device Identifier10886982068835
UDI-Public(01)10886982068835
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K111667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.010.441
Device Catalogue Number03.010.441
Device Lot Number180064-103
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Initial Date Received by Manufacturer 11/30/2020
Supplement Date Received by Manufacturer01/12/2021
Initial Report FDA Received Date12/28/2020
Supplement Report FDA Received Date01/19/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
11MM TI CANNULATED TIBIAL NAIL-EX/375MM-STERILE; 12.0MM/8.0MM PROTECTION SLEEVE 188MM; 8.0MM/4.2MM DRILL SLEEVE 200MM; INSERTION HANDLE FOR SUPRAPATELLAR; UNK - NAILS; 11MM TI CANNULATED TIBIAL NAIL-EX/375MM-STERILE; 12.0MM/8.0MM PROTECTION SLEEVE 188MM; 8.0MM/4.2MM DRILL SLEEVE 200MM; INSERTION HANDLE FOR SUPRAPATELLAR
Patient Age29 YR
-
-