| Model Number |
2P56-22 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complete information for section a patient information, 1.Patient identifier = sid= (b)(4) an evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely elevated lactate dehydrogenase (ldh) results on architect c16000 processing module for one patient.The results provided were: on (b)(6) 2020 sid (b)(6) initial ldh= 371 u/l /repeated=582 u/l /repeated=366 u/l and 380 u/l.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of complaints found no trends for the complaint issue.Return testing was not completed as returns were not available.No file sample testing performed as the issue is sample / tube specific at the customer site; the customer generated acceptable results using the same reagent.Device history record review did not identify any non-conformances or deviations with the likely cause lot and complaint issue.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.Labeling was reviewed and multiple probable causes were identified: handle specimen collection and sample tube processing per the tube manufacturer's instructions.So handled, plasma may still contain cellular elements which can interfere with the ldh measurements due to presence of the enzyme in the cells or cellular debris or sample handling issues.Based on the investigation no product deficiency was identified for the lactate dehydrogenase (ldh) assay reagent, lot 02747un20.
|
| |
|
Search Alerts/Recalls
|