• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR TRABECULAR METAL POSTERIOR STABILIZED STANDARD POROUS; PROSTHESIS KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR TRABECULAR METAL POSTERIOR STABILIZED STANDARD POROUS; PROSTHESIS KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Reaction (2414)
Date of Event 11/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device whereabouts are unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products and mdrs: 42521400811 articular surface fixed bearing posterior stabilized right 11 mm lot# 63378992 mdr: 0002648920-2020-00541.42532007502 tibia cemented 5 degree stemmed right size f lot# 63603170 mdr: 3007963827-2020-00336.
 
Event or Problem Description
Patient called stating he underwent a right tka.Subsequently, approximately 3 years post procedure the patient was revised due to poly failure causing metal debris in the knee.Patient is seeking compensation.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: therapy performed due to pain, instability, weakness, use of assistive device, performed thereafter for left tka and right knee ongoing pain, patient has 3/5 mmt for knee extension/flexion; patient has crepitus around knee, limp right leg, failed conservative therapy treatment, adhesion/scar tissue removal, instability and pain that implant adl, very thickened black synovium consistent with metallosis, metal stained synovium, tibial insert had disengaged from tray, burnishing of posterior condyles consistent with metal wear, tibial tray showed significant wear with locking mechanism destroyed, removed femoral and tibial components, debridement of metal stained synovium, patella and patella component well fixed and tracked well.Complaint is confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR TRABECULAR METAL POSTERIOR STABILIZED STANDARD POROUS
Common Device Name
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key11083809
Report Number3007963827-2020-00337
Device Sequence Number14391257
Product Code JWH
UDI-Device Identifier00889024469945
UDI-Public(01)00889024469945
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K122745
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42500806802
Device Lot Number63766860
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/01/2020
Supplement Date Received by Manufacturer02/17/2021
Initial Report FDA Received Date12/28/2020
Supplement Report FDA Received Date02/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight116
-
-