| Model Number |
1013317J |
| Medical Device Problem Code |
Component Misassembled (4004)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a lesion in the moderately calcified, moderately tortuous left anterior descending coronary artery.During use, an unknown marker was found on a versaturn f guide wire.Unspecified devices were used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual analysis was performed on the returned device.The component misassembled / extra marker was not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined the cause for the reported unknown / extra marker is due to the solder used to manufacture the device, that was likely mistaken for an extra marker.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|