• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS TORIC II 1-PIECE ACRYLIC; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS TORIC II 1-PIECE ACRYLIC; TORIC IOLS Back to Search Results
Model Number ZCU375
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided.Best estimate of date of event is between 6/8/2020 and 11/23/2020.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was implanted in the patient¿s left eye, however five months post-op refraction was -1.25 + 2.00 x 90, and lens was placed at 170.Pre-op astigmatism was 2.06 x 170.Patient is not post-lasik and the iol is in perfect position (no rotation) post-operatively.It was noted that the issue is post-op refractive surprise.There was no anatomy or quality issues.Upon further follow-up, it was confirmed that the iol was later explanted.No other information was provided.
 
Additional Manufacturer Narrative
Additional information: section d10: device available for evaluation? yes.Returned to manufacturer on: 1/25/2021.Section h3: device returned to manufacturer? yes.Device evaluation: the complaint lens was received inside a specimen cup, with customer notes.Visual inspection under magnification, revealed viscoelastic residue on the optic body and haptics.Which is consistent with a lens that was handled, during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed.And no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed, that the product was manufactured and released according to specifications.A search revealed, that no other complaint was received from this production order.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS TORIC II 1-PIECE ACRYLIC
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key11085611
Report Number9614546-2020-00542
Device Sequence Number16894100
Product Code MJP
UDI-Device Identifier05050474700468
UDI-Public(01)05050474700468(17)241210
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/10/2024
Device Model NumberZCU375
Device Catalogue NumberZCU375U150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2021
Initial Date Received by Manufacturer 12/02/2020
Supplement Date Received by Manufacturer02/14/2021
Initial Report FDA Received Date12/29/2020
Supplement Report FDA Received Date02/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-