| Model Number |
VENUM14060 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Expiry date (07/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement through right common femoral vein, the device allegedly failed to expand in the distal end.It was further reported that the physician jiggled to deploy the stent.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the sample has been returned to the manufacturer for evaluation.The condition of the returned catheter sample confirmed the alleged expansion issue.The silicone brake of the inner catheter which is under the stent was found shifted to distal which indicated that the stent adhered to the brake leading to breakaway and shifting upon removal.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of the relevant labeling for this product was conducted.The instruction for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: b5, d4 (expiry date: 07/2022), g3 h11: h6 (method, result) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement in the right iliac vein through ipsilateral popliteal access, the tail end of the stent allegedly failed to expand.It was further reported that the physician jiggled the system to finally deploy the stent.There was no reported patient injury.
|
| |
|