• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. VANTAGE; TIBIAL PLATE FB SZ 3 RT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. VANTAGE; TIBIAL PLATE FB SZ 3 RT Back to Search Results
Model Number 350-12-03
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The loosening reported was likely the result of either an insufficient bond between the implant and the bone, patient-related conditions related to medical history and/or previous operations on involved ankle, or a combination of the two.However, this cannot be confirmed as the patient has not yet been revised; therefore, the devices were not available for evaluation.Concomitant device(s): 350-22-62, tibial insert fb sz 2 rt 12mm, 350-02-01, talar implant sz 1 rt.
 
Event or Problem Description
As reported, the (b)(6) male patient presented with pain "like before my surgery" after beginning a walking program.Patient¿s bmi is (b)(6), weight is (b)(6).X-rays show cyst formation and early loosening in right ankle.The patient was instructed to reduce walking and return in 2 months.Patient has a history of post traumatic arthritis, hypertension, a-fib; pad; h/o pulmonary embolism & skin cancer.The case report form indicates this event is possibly related to devices or procedure.This event report was received through clinical data collection activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VANTAGE
Common Device Name
TIBIAL PLATE FB SZ 3 RT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key11086722
Report Number1038671-2020-00670
Device Sequence Number11475450
Product Code HSN
UDI-Device Identifier10885862276193
UDI-Public10885862276193
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152217
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number350-12-03
Device Catalogue Number350-12-03
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/16/2020
Initial Report FDA Received Date12/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient Weight109
-
-