• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII OVAL RESURFACING PAT 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII OVAL RESURFACING PAT 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71421029
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Joint Laxity (4526)
Date of Event 12/04/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after tka, patient presented pain.A ct showed hot spots and loosening on tibial component and patella.A total knee revision surgery was performed to implant competitor's products.Patient outcome unknown.
 
Additional Manufacturer Narrative
It was reported that after tka, patient presented pain and a total knee revision surgery was performed.Ct showed hot spots and loosening on tibial component and patella.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident or adverse event could not be corroborated.A clinical evaluation noted that no consent has been given to provide the requested surgical reports and x-rays.Therefore, the patient impact beyond the revision cannot be determined.Due to the limited information provided, no further clinical assessment is warranted.Based on this investigation, the need for corrective action is not indicated.Some potential causes could include but are not limited to fit/sizing, traumatic injury, abnormal motion over time or patient medical history.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII OVAL RESURFACING PAT 29MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11086813
Report Number1020279-2020-07863
Device Sequence Number19660866
Product Code JWH
UDI-Device Identifier03596010485052
UDI-Public03596010485052
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/26/2026
Device Model Number71421029
Device Catalogue Number71421029
Device Lot Number16AM19174
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/04/2020
Supplement Date Received by Manufacturer05/15/2021
Initial Report FDA Received Date12/29/2020
Supplement Report FDA Received Date05/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
-
-