Model Number 1550350-15
Health Effect - Clinical Codes Perforation (2001); Stenosis (2263)
Date of Event 12/01/2020
Type of Reportable Event Malfunction
Event or Problem Description
Three devices failed.Incident description: opened, prepped, and advanced three stent catheters.Unable to advance to lesion site in the mid circumflex artery.Removed over wire and all three stents had damage to the stent struts.There were no complications.From the procedural note: intervention: for purpose of intervention, i ultimately upsized to a 7-french system for complex bifurcation intervention.I initially started with a 6-french, and crossed into the first marginal and into the distal av circumflex, after achieving an act (activated clotting time) greater than 300 seconds with a ____ and a pilot wire.A predilation was initially performed into the main av groove circumflex with a 3.0x12 mm balloon.Initial plan given the anatomic configuration was to perform simultaneous stenting with a "pair of pants" anatomic configuration.Unfortunately, given that the 2 wires repetitively became twisted and on top of each other, it did not allow passage of the second stent down the vessel.I had to utilize several different stents and repositioned wires on multiple occasions to attempt to address the issue with switching out, which then we delivered first into the av groove circumflex, the first marginal and neither were successful.I upsized to a 7-french system and additionally tried the same thing with a 7-french cls 3.5 guiding catheter to assess whether there it was guide catheter size related issue.The same issue was involved with wires twisting upon each other.At this point, several stents were unable to pass for this reason, as they could not pass through the guiding catheter, even despite upsizing to a 7-french system.At this point, i elected to simplify the procedure and proceed with stenting of the main branch and balloon angioplasty of the side branch.At this point, utilizing 2 pilot wires, which were placed distally, i stented the main branch with a 3.5x18 mm xience alpine sierra drug-eluting prosthesis, and then re-crossed with a pilot wire into the first marginal branch.I then performed simultaneous kissing balloon angioplasty with a 2.75 mm balloon in the side branch with a 4.0 mm balloon in the main branch.The final result was excellent.There was brisk flow into the av groove circumflex with kissing balloon angioplasty.There was a residual, less than 50% stenosis within the first obtuse marginal.Despite multiple passes of balloon, stents, wires, and different guiding catheter, the left main did not appear different at the end of procedure, and there was no evidence of any distal perforation or distal compromise, despite extensive wire exchanges.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
ABBOTT VASCULAR INC.
26531 ynez road
temecula CA 92591
MDR Report Key11087020
Report Number11087020
Device Sequence Number3692066
Product Code NIQ
UDI-Device Identifier08717648227417
UDI-Public(01)08717648227417(17)220116(10)0070242(91)5716
Combination Product (Y/N)N
Initial Reporter StateFL
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 12/10/2020
Report Date (Section F) 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number1550350-15
Device Catalogue Number1550350-15
Device Lot Number0070242
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Event Location Hospital
Date Report to Manufacturer12/29/2020
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight134
Date Report Sent to FDA12/10/2020
-
-